| Temperature | Set temperature range in °C | min. 5°C above ambient up to +300°C |
| | Temperature | 1 Pt100 sensor DIN class A in 4-wire-circuit |
| Control technology | Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
| | Control COCKPIT | SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display |
| | Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
| | Function SetpointWAIT | the process time does not start until the set temperature is reached |
| | Calibration | three freely selectable temperature values |
| | Adjustable parameters | temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime |
| Ventilation | Convection | natural convection |
| | Fresh air admixture | adjustment of pre-heated fresh air admixture by air flap control in 10 % steps |
| | Vent | vent connection with restrictor flap |
| Communication | Documentation | programme stored in case of power failure |
| | Programming | AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand). |
| Safety | Temperature control | adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
| | Autodiagnostic system | for fault analysis |
| Standard equipment | Works calibration certificate | Calibration at +160°C |
| | Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
| | Internals | 1 stainless steel grid |